Infant Transport Incubator

Transport Incubator
  • Two temperature control modes:  Air mode and baby mode controlled by micro-computer
  • Use AC Power and DC Power alternatively, DC 12V or DC 24V Power on the ambulance
  • Set temperature, air temperature, baby temperature and internal battery voltage can be displayed separately
  • Failure alarm indication
  • Can be equipped with >.7 deg C temperature set function
  • Double wall hood with side door so that the bassinet can be pulled out from it
  • Natural air flow humidity
  • The height of whole unit can be adjusted
  • Oxygen supply system
  • Observation lamp
Standard Configuration
Main body (including the hood, controller, bassinet, internal battery, and observation lamp,  oxygen supply system, skin temperature sensor, I.V. Pole, mattress, adjustable stand)
Optional configuration: >37 deg C override temperature mode
Specification
  • AC Power supply:  AC220V – 230V/50Hz
  • DC Power supply:  DC12V/10A or DC24V/6A
  • Input power:  400VA
  • Mode of control:  Air mode and baby mode controlled by micro-computer
  • Air temperature control range:  25 deg C – 37 deg C

(Override mode:  37 deg C – 38 deg C)

  • Baby temperature control range:  34 deg C – 37 deg C

(Override mode:  37 deg C – 37.5 deg C)

  • Skin temperature sensor accuracy:  <0.3 deg C
  • Temperature variability:  <1.0 deg C
  • Temperature uniformity:  <1.5 deg C (mattress is on the horizontal position)
  • Noise inside hood:  <55dB(A) (Environment noise <45dB(A))
  • Failure alarm:  Over temperature alarm, temperature deviation alarm, sensor alarm, motor fan failed alarm, power failure alarm and so on
  • Operating at temperature equilibrium:  90min (Set temp 36 deg C, Ambient 15 deg C)
Environmental Temperature
Operating range:  10 deg C – 30 deg C

(Set temperature should be higher 3 deg C than the environment temperature)

Ambient air movement rate:  <1.0m/s