YP-90 Series Infant Incubator

Infant Incubator
  • Two control modes:  Air mode and baby mode controlled by micro-computer
  • Set temperature, air temperature, baby temperature are displayed separately
  • Natural air flow humidity
  • Inclination of bassinet is adjustable
  • Failure alarm indication
  • RS-232 connector
  • YP-90AB is configured with fixed Neonate Bilirubin Phototherapy Equipment:  (with blue fluorescent lamps)
  • YP-90AB is configured with fixed Neonate Bilirubin Phototherapy Equipment:  (with LED bulbs) as optional
Standard Configuration
YP-90A:  The hood, the base, the mobile cabinet, the controller, I.V. pole, shelf
YP-90AB:  Based on YP-90A, with Neonate bilirubin phototherapy equipment (fixed model)
Optional parts: >37 deg C temperature override Mode function, vertical Height Adjustment Stand (VHA stand)
Specification
  • Power supply: AC220V-230V/50Hz
  • Max Power Output:  850VA
  • Mode of control:  Air mode and baby mode controlled by micro-computer
  • Air temperature control range: 25 deg C – 37 deg C

(Override mode: 37 deg C – 39 deg C)

  • Baby temperature control range: 34 deg C – 37 deg C

(Override mode: 37 deg C – 38 deg C)

  • Accuracy of skin temperature sensor:  <0.3 deg C
  • Temperature variability:  <0.5 deg C
  • Temperature uniformity:  <0.8 deg C (mattress is on the horizontal position)
  • Inclination of bassinet:  +/- 12 deg
  • Noise inside hood:  <55dB(A) (Environmental noise <45dB(A))
  • Failure alarm:  Over temperature alarm, deviation alarm, sensor failure alarm, fan motor alarm, power failure alarm, system alarm and so on
Environmental Temperature
Operating range:  20 deg C – 30 deg C
(Set temperature should be higher 3 deg C than the environment temperature)
Ambient air movement rate:  <0.3m/s