ECG1200G Digital 12-channel ECG

 

ECG1200G Digital Twelve Channel ECG Patient Monitor
ECG1200G 12-channel ECG gathers 12-lead ECG signals simultaneously and prints out the ECG waveforms with thermal printing system.  Its users are mainly hospitals and all kinds of medical establishments.
Technical Specifications
Input Way Floating and defibrillation protection
Lead Standard 12 leads
Patient Leak Current <10 micro A
Input Impedance >50M Ohm
Frequency Response 0.05Hz – 150Hz (-3dB)
Time Constant Time constant >3.2s
CMRR >60dB, >100dB (Adding filter)
EMG interference filter 35Hz(-3dB)
Recording Way Thermal printing system
Specification of Recording Paper 210mm (W) x 20m (L) x high-speed thermal paper
Paper Speed Auto record: 25mm/s, 50mm/s, error +/-5%
Rhythm record: 25mm/s, 50mm/s, error +/-5%
Manual record: 5mm/s, 6.25mm/s, 10mm/s, 12.5mm/s, 25mm/s, 50mm/s, error +/-5%
Sensitivity Selections 5, 10, 20, 40mm/mV, error +/-5%, Standard sensitivity is 10mm/mV +/-0.2mm/mV
Auto Record Record setup according to auto record format and mode, automatically changing leads, measuring and analysing
Rhythm Record Record setup according to rhythm record format and mode, automatically measuring and analysing
Manual Record Record setup according to manual record format, manually changing leads
Measurement Parameters HR, P-R interval, P Duration, QRS Duration, T Duration, Q-T Interval, Q-Tc, P Axis, QRS Axis, R(V5), S(V1), R(V5) + S(V1)
Product Safety Type Class 1 CF applied part; there is defibrillation and pacing protection circuit in it
Enduring Polarization Voltage +/-500mV
Noise Level <15 micro Vp-p
Size 334mm(L)x320mm(W)x85mm(H)
Power Supply/Frequency AC100-240V, 50/60Hz, DC 14.8V, 3700mAh, rechargeable lithium battery