Advance-150 Multi-Parameter Patient Monitor

Advance-150 Multi-Parameter Patient Monitor
  • 15” high resolution colour TFT screen.
  • 5 or 9-channel waveform display with trend coexistence, switchable to large font.
  • Monitoring standard parameters of ECG, SpO2, RESP, NIBP, TEMP & pulse.
  • Automatic analysis of 100 arrhythmia waveforms.
  • S-T segment with automatic & manual analysis.
  • Digital SpO2, technology with anti-motion & anti-low infusion.
  • Trend data analysis for 8/24/120 hours.
  • Storage & review 500 groups of data & 24-hour ECG waveforms.
  • Protection against interference from defibrillator & electrosurgical unit.
  • Feasible to be used with cardiac pacemaker together.
  • Optimized models for adult, child & neonate.
  • Removable rechargeable battery.
  • Network capability (optional).
TECHNICAL SPECIFICATION
ECG  
ECG Amplifier Bandwidth Monitoring mode: 0.5-40Hz

Diagnosis mode: 0.05-75Hz

CMRR >80dB
Heart

Rate

Range 20-250bpm
Accuracy +/-2bpm

ECG Gain

1/2

5mm/mV +/- 5%

1

10mm/mV +/- 5%

2

20mm/mV +/- 5%

Auto

Auto gain control
Lead Selection GND: Standard I; II; III; aVR, aVL, aVF; chest V1–V6
ECG Sweeping Speed 12.5mm/s, 25mm/s, 50mm/s
Alarm Setting Customized upper & lower limits & autosaved
Sp02
Display Sp02 value & waveform, pulse rate & histogram
Measurement Range 0-99%
Resolution +/-1%
Accuracy of Sp02 +/- 2% (75-99%)
+/- 3% (50-75%)
Measurement Range of Pulse Rate 30-240 bpm
Accuracy of Pulse rate +/- 2bpm
Alarm Setting Customized upper & lower limits & autosaved
RESP
Measure Method Thoracic impedance
Measure Range 0-100 bpm
Accuracy +/- 5%
Alarm Setting Customized upper & lower limits autosaved
TEMP
Range 28.0 – 45.0 deg C
Accuracy +/- 0.2 deg C
Alarm Customized upper & lower limits & autosaved
NIBP
Measure Method Intelligent oscillometric method
Parameters Systolic, diastolic & mean pressure
Working Modes Manual, automatic & STAT (5 minutes)
Unit Display mmHg/kPa selectable
Cuff Pressure Range 0-300 mmHg
Cuff Type Adult (standard)

Child/neonate (optional)

Automatic Measuring Time 1-240 minutes
Inflation Time <10s (typical adult cuff)
Deflation Time <2s (typical adult cuff)
Monitoring

Range

Adult/Child Systolic 25-265mmHg
Diastolic 12-220mmHg
Mean 18-245mmHg
Neonate Systolic 25-135mmHg
Diastolic 12-110mmHg
Mean 18-120mmHg
Alarm Setting Customized upper & lower limits & autosaved
General Specifications
Standard Configuration ECG, HR, RESP, NIBP, Sp02, PR, TEMP, Battery
Power 100-250VAC, 50/60Hz
Power Consumption <80VA
Display TFT Colour LCD (15”)
Alarm Audible & visual
Networking RJ45
Battery Rechargeable lead-acid battery
Service Life >2 hours
Safety Standard EN IEC60601-1 (GB9706.1)
Optional Module Built-in thermal printer
1-channel IBP
2-channel IBP
Sidestream CO2
Mainstream CO2